Preview

Ac Hipot Verification Procedure.

Satisfactory Essays
Open Document
Open Document
403 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Ac Hipot Verification Procedure.
AC Hipot Verification procedure.

The AC Hipot test checks the insulation of a device to make sure that there is not a voltage breakdown between the electrical components and conductors in the device, and the chassis ground of the device.

A high AC voltage is applied to the connection wires of the device, and the frame of the device is grounded. The return side of the hipot unit must also be connected to ground. In some cases, it is necessary to float the return and not connect it to ground, but that will not be considered in this writeup.

The hipot can fail for having high leakage current, a ground, an ARC, or low leakage current.

To verify that the hipot test is working, do the following...

First, disable all other tests except hipot.

1. To verify that high leakage is detected, temporarily adjust the hipot maximum, either in the hipot itself, or in the software, lower than the actual, and see if it fails for high leakage.

2. To verify that a ground is detected, skip all other tests but hipot, and place a wire between the test clip and the chassis ground of the device. The test should fail for a ground. A high leakage reject is also acceptible in this exercise.

3. To verify that an ARC is detected, connect a wire to chassis ground. The other end, hold in your hand and wait. Start a hipot test, after the hipot test has started, and before it is complete, touch the other end of the wire to clip 1 of the clip block, or to a conductor on the test connector being used. The test should fail for ARC.

4. To verify that low leakage is detected, first, completely disconnect the device from the clip block, or test connector. The test should fail. Then, reconnect the device under test to the clip block, or test connector, and isolate the device from where it is connected to the chassis of the machine. If the device sits in a metal fixture, then set the device on delrin, or wood. If the device has a ground wire, disconnect and

You May Also Find These Documents Helpful

  • Satisfactory Essays

    This is an item of electrical test equipment used to determine if an electrical path can be established between two points; that is if an electrical circuit can be made. The circuit under test is completely de-energized prior to connecting the tester. The tester consists of an indicator in series with a source of electrical power - normally a battery, terminating in two test leads. If a complete circuit is established between the test-leads, the indicator is activated (the indicator may be an electric light or a buzzer).…

    • 256 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Flvs Chem 04.05 Lab

    • 571 Words
    • 3 Pages

    Use a U-tube with electrodes at each end, connected to a battery. Fill the U-tube with water. Turning on the battery, observe the results at each electrode. Next, test for hydrogen and oxygen gas produced at each end of the tube.…

    • 571 Words
    • 3 Pages
    Good Essays
  • Powerful Essays

    Est1 Task 5

    • 3478 Words
    • 14 Pages

    hub centre line & bending them simultaneously inward & then outward & check for abnormal noise.…

    • 3478 Words
    • 14 Pages
    Powerful Essays
  • Good Essays

    Lab 18

    • 846 Words
    • 3 Pages

    B. Touching the positive conductor should result in a reading that is the same as the battery output…

    • 846 Words
    • 3 Pages
    Good Essays
  • Satisfactory Essays

    repeatedly with no problems. You can customize the tests to bring out hidden information from the…

    • 442 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Schedule a test to verify the peripheral are tested to ensure the correct peripheral devices is being plug in and ensure usb drive/cd rom are disable.…

    • 960 Words
    • 4 Pages
    Good Essays
  • Good Essays

    Brownfields

    • 547 Words
    • 3 Pages

    2. Designate ten spots for testing for contaminants; these spots should be located at five-foot intervals between the leaking underground storage tanks and Boston Post Road.…

    • 547 Words
    • 3 Pages
    Good Essays
  • Satisfactory Essays

    lab 24

    • 332 Words
    • 2 Pages

    Perform the molisch, benedict, and iodine tests. Perform a grease spot test and then do the xanthoprotic, biuret, and ninhydrin tests.…

    • 332 Words
    • 2 Pages
    Satisfactory Essays
  • Satisfactory Essays

    Corn and Milk Lab

    • 290 Words
    • 2 Pages

    6)Use a clean pipet for each indicator,add 15 drops of each indicator to the test tubes…

    • 290 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Chemistry

    • 390 Words
    • 2 Pages

    Use a U-tube with electrodes at each end, connected to a battery. Fill the U-tube with water. Turning on the battery, observe the results at each electrode. Next, test for hydrogen and oxygen gas produced at each end of the tube.…

    • 390 Words
    • 2 Pages
    Good Essays
  • Good Essays

    Circuits 2014

    • 604 Words
    • 3 Pages

    Click in the Tools boxes to get a voltmeter and a Non-contact ammeter. Put the voltmeter near the battery and place the red tab at one end and the black at the other…

    • 604 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Diffusion is the process by which collisions between molecules cause to spread apart. This movement is described as movement from an area of greater concentration to an area of lower concentration. Hence, diffusion continues until the molecules are equally distributed. This is to ensure that molecules have reached a state of equilibrium. Diffusion occurs spontaneously, no energy is involved.…

    • 2007 Words
    • 9 Pages
    Good Essays
  • Better Essays

    - Second, add the materials to be tested into each tube in the right order and…

    • 1405 Words
    • 16 Pages
    Better Essays
  • Good Essays

    Bs&W Procedure

    • 997 Words
    • 4 Pages

    5. Open the sampling point and flush the sample line into an earthed waste container for as long as needed to ensure the pipe work is hot enough.…

    • 997 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    Mechanic

    • 275 Words
    • 2 Pages

    Print names and enter signatures and dates to certify that the persons named in this…

    • 275 Words
    • 2 Pages
    Satisfactory Essays

Related Topics