Preview

Fluid Flow in a Duct of Varying Cross-Section Report

Powerful Essays
Open Document
Open Document
1910 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Fluid Flow in a Duct of Varying Cross-Section Report
City University London

Fluid Flow in a Duct of Varying Cross-Section Report:

Khurshidanjum Pathan, Group A1a

Abstract:

The experiment is carried out to demonstrate the relation between pressure and fluid velocity in a duct of varying cross-section by using Bernoulli’s equation and continuity equation.(1) Bernoulli’s equation relates the pressure to the velocity for a fluid of constant density flowing in a Venturi tube. Static head, normalised head and percentage of errors were calculated using the result of the data. On the base of calculation its being analysed that in the contraction flow, velocity and dynamic head increases with decreased static pressure and dynamic pressure. While in the convergent flow, velocity decreases with increased static pressure and dynamic pressure.

Introduction:

[pic]

Figure (1) sketch of Apparatus used to investigate the validity of Bernoulli equation (3)

[pic]

Figure (2) Apparatus used to investigate the validity of Bernoulli’s equation

[pic][pic]

Figure (3) Water pump Figure (4) Venturi pipe

[pic][pic]

Figure (5) Manometer tubes Figure (6) Flow meter

Figure (1)-(2) shows the apparatus used for the experiment consist of a acrylic duct of varying cross-section known as a Venturi as shown in Figure (4). The duct has 10 wall tappings for the measurement of static pressure in both convergent and divergent area of a duct. This duct is connected to a flow meter on its right side, where the flow rate is being set as shown in the figure (6). On its left side apparatus is being connected to manometer through 10 different pipes and readings at each of the tapping for each flow rate are being recorded as shown in figure (5). Pressure of water is being controlled by attached hand pump by adjusting valve as shown in figure (3)

| | |



References: (3)http://www.scribd.com/doc/68038037/Experiment-Bernoulli Richard A Furness, Fluid Flow Measurement, Longman, 1989, page 7-11-12 B S Massey, Mechanics of Fluids, Sixth edition, 1989, page 76-77-78-79 Appendices:

You May Also Find These Documents Helpful

  • Satisfactory Essays

    For ‘EXPERIMENTAL’ tubes – state the purpose, results AND interpret the results of the experiment.…

    • 592 Words
    • 3 Pages
    Satisfactory Essays
  • Better Essays

    The objective of this exercise is to measure the pressure distribution across the surface on an aerofoil in a wind tunnel. The aerofoil is tested under several different Mach numbers from subsonic to supercritical. The purpose of measuring the pressure distributions is to assess the validity of the Prandtl-Glauert law and to discuss the changing chracteristics of the flow as the Mach number increases from subsonic to transonic.…

    • 2069 Words
    • 9 Pages
    Better Essays
  • Good Essays

    First we will set up the experiment as shown in Figure 2, and then we will fill the trough with room temperature water and record the temperature and we will also add the room temperature water to the buret until it is about 10ml from the top. If there are no leaks in the apparatus, we will add 10ml of KI to 15ml of distilled water in a 125ml flask. We will swirl the flask so that it achieves room temperature, then we will add 5ml of H2O2 and quickly stopper the flask. While one student continues to swirl the flask, the other should observe the volume of oxygen produced, after 2ml of gas has been evolved. The recordings should be written down at 2ml intervals until 14ml of oxygen has been evolved. Next, we will repeat the experiment (making sure everything is the same, including the temperature of the bath water) by adding 10ml of KI and 10ml of H2O2 (after swirling the flask), then quickly stoppering the flask and measuring the readings. This will all be repeated again for a third time by adding 20ml of KI and 5ml of distilled water, swirling and then adding 5ml of H2O2.…

    • 472 Words
    • 2 Pages
    Good Essays
  • Powerful Essays

    Unit two Biology

    • 7492 Words
    • 30 Pages

    As the two tubes are connected by the manometer these pressure changes are cancelled out. Pressure changes affecting one tube but not the other will cause the manometer fluid to show a difference in height on the two sides of the U-tube. The apparatus is assembled with both three-way taps open to the air, to prevent the manometer fluid being pushed into the tubes. However, during an experiment both three-way taps are closed to prevent air movement between the air in the apparatus and air in the laboratory, as shown in Figure 2.18. The difference between the levels of manometer fluid on the two sides of the U-tube represents the pressure difference between the air in the two tubes.…

    • 7492 Words
    • 30 Pages
    Powerful Essays
  • Good Essays

    Ap Biology Enzyme

    • 425 Words
    • 2 Pages

    3) The experimental system of this lab is, first to set up the logger pro pressure measuring device to the computer. After the device is set up completely, set up the logger pro software that will collect the data for this lab. After both of these steps are complete, get a clean test tube with 1.5 mL of H2O, 1.5mL of H2O2 and then add drops of enzyme needed. Collect data by plugging the black “plug” into the test tube and press “start” on the logger pro software. Continue collecting data with variety of different conditions.…

    • 425 Words
    • 2 Pages
    Good Essays
  • Good Essays

    Microscale Lab

    • 709 Words
    • 3 Pages

    ABSTRACT: The objective of the experiment was to be introduced to working with the lab equipment, specifically heating and measuring, and to be able to create some of the apparatus that future experiments will require. Heating methods used were an aluminum block, a sand bath, and a water bath. The aluminum block heated most quickly while the others were more stable. Regarding the measurement of water, the less volume of water that was being transferred, the less accurate the measurement turned out to be.…

    • 709 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Physiology Answers

    • 3429 Words
    • 15 Pages

    A.the flow of fluid from point A to point B in a tube depends upon the difference in pressure between A and B.…

    • 3429 Words
    • 15 Pages
    Good Essays
  • Good Essays

    Artificial Breathing Aids

    • 695 Words
    • 3 Pages

    The patient's head was positioned outside the box while the rest of the body was enclosed in the airtight metal box. A pump connected to the box varied the air pressure inside the chamber. When the air pressure inside the box decreased, the weight of the atmosphere outside the box forced air through the nose and mouth into the lungs. When the air pressure in the box increased, the air was forced out of the lungs.…

    • 695 Words
    • 3 Pages
    Good Essays
  • Powerful Essays

    4. By releasing the tap gradually or pressing the pump harder, the reading of the…

    • 2729 Words
    • 11 Pages
    Powerful Essays
  • Powerful Essays

    experiment is done to investigate the validation of the Bernoulli’s equation and also to measure pressure distribution along venture tube.…

    • 890 Words
    • 4 Pages
    Powerful Essays
  • Powerful Essays

    venturi meter

    • 914 Words
    • 4 Pages

    pressure difference between the venturi inlet and neck, and from this the flow-rate can be…

    • 914 Words
    • 4 Pages
    Powerful Essays
  • Satisfactory Essays

    Bernoullis Apparatus

    • 258 Words
    • 2 Pages

    when appliedto the steady flow of water in tapered duct and to measure the flow rates and both static andtotal pressure heads in a rigid convergent/divergent tube of known geometry for a range of steady flow rates. The…

    • 258 Words
    • 2 Pages
    Satisfactory Essays
  • Satisfactory Essays

    4) The pressure reading inside the chamber was monitored until it stabilizes. The pressure reading P1 and temperature T1 was recorded.…

    • 651 Words
    • 3 Pages
    Satisfactory Essays
  • Good Essays

    Prosti

    • 721 Words
    • 3 Pages

    The pitot-static system of instruments uses the principle of air pressure gradient. It works by measuring pressures or pressure differences and using these values to assess the speed and altitude. These pressures can be measured either from the static port (static pressure) or the pitot tube (pitot pressure). The static pressure is used in all measurements, while the pitot pressure is only used to determine airspeed.…

    • 721 Words
    • 3 Pages
    Good Essays
  • Better Essays

    Friction Loss Along Pipe

    • 2637 Words
    • 11 Pages

    This experiment of the friction loss along a smooth pipe shows that there are existence of laminar and transitional flows as stated in Graph 2.0 and Graph 2.1. It is proven that the higher velocity along the smooth bore pipe, the higher is the head loss of water. As shown in Table 3.0, when the Reynolds’ number increases, the value of pipe coefficient friction, f decreases along the decreasing stead laminar line. On top of that, there are energy loss from the water to the surface of the pipe and therefore, the temperature increases when velocity, flow rate and head loss increases respectively. The percentage difference of obtained head loss and calculated head loss are 2.5%, 19.0%, 32.0%, 27.0% and 30.0% whereby the differences are not major and in the acceptable range. There are few factors in affecting the head loss which are flow rate, inner diameter of the pipe, roughness of the pipe wall, corrosion and scale deposits, viscosity of the liquid, fittings and also straightness of the pipe. There are existence of both human errors, parallax errors and environmental effect but there are always error counters to be taken place to increase the accuracy of the results.…

    • 2637 Words
    • 11 Pages
    Better Essays

Related Topics