The calibration with Hidex 300 SL was also performed using the two previously mentioned ways. Using the manufacturer' standard solution, the efficiency achieved was 98 and 73% for alpha and beta counting, respectively. Whereas, using the internal (mixed 214Am-90Sr) standard source the Pulse Length Index (PLI) to discrimination of alpha and beta at 6. The triple-to-double coincidence ratio (TDRC), which the same as the efficiency, was 100% for alpha and 99.75% for beta counting. To set window for counting 226Ra, a standard source of 226Ra activity of 2 Bq mL-1 was made in lab. The net count of total area under window from 550 to 750 with TDCR of 300% was used because this window covers the 222Rn, 218Po and 214Po
The calibration with Hidex 300 SL was also performed using the two previously mentioned ways. Using the manufacturer' standard solution, the efficiency achieved was 98 and 73% for alpha and beta counting, respectively. Whereas, using the internal (mixed 214Am-90Sr) standard source the Pulse Length Index (PLI) to discrimination of alpha and beta at 6. The triple-to-double coincidence ratio (TDRC), which the same as the efficiency, was 100% for alpha and 99.75% for beta counting. To set window for counting 226Ra, a standard source of 226Ra activity of 2 Bq mL-1 was made in lab. The net count of total area under window from 550 to 750 with TDCR of 300% was used because this window covers the 222Rn, 218Po and 214Po