When recording data, each entry should be given a corresponding estimated error, or uncertainty. The uncertainty gives the reader an idea of the precision and accuracy of your measurements. Use the following method for finding the uncertainty associated with any measuring device used in lab.
First, find the least count, or the smallest printed increment, of the measuring device. On the meter sticks, the least count is 1 mm. On the double pan balances, the least count is 0.1 g. On the small graduated cylinders, the least count is 25 ml. If you are using the full precision of the instrument, you are probably safe in saying that your measurement is within one least count of the measured value, in either direction.
Figure 1
For example, say you are measuring the object in Figure 1. If you use the meter stick to measure an object's length as being around 86 mm, that means you are pretty sure that the actual value is between 85 mm and 87 mm. Therefore, you should represent your data like this: l = 86 mm ± 1 mm
In this case, your uncertainty is ± 1 mm. However, you may feel that you are able to attain more precision than is indicated by the least count. In that case, you should do some estimating. By estimating, you divide the least count of your measuring device into imaginary increments. In this lab, it is recommended that you divide the least count into five imaginary increments. This is called the one-fifth rule. There will be some occasions when the one-fifth rule seems too generous. If you feel that your confidence in the last significant figure of the measurement is greater than this, then of course it would be more appropriate to use, say one-tenth of the least count. Similarly, if your confidence in the last significant figure is lower, then you might use half the least count. At any rate, you should use good judgment in estimating the error. Always think in terms of having to justify your estimates to your instructor! The following data points were estimated to the nearest half of the least count and the nearest fifth of the least count, respectively. l = 85.5 mm ± .5 mm l = 85.6 mm ± .2 mm
All three of these length measurements are correct, but they represent varying degrees of precision. In each case, the uncertainty was decreased, indicating greater accuracy of measurement. You should always try to be as precise as possible, and by estimating, you were able to attain a third significant digit (see appendix B for more about significant digits).
There are two rules to follow to help make sure you have determined your uncertainty correctly. Note that these rules were followed in each case above. * Rule 1: The uncertainty should be to the same precision as the measured value. * Rule 2: The measured value should be evenly divisible by the uncertainty.
In the case of a digital device, the uncertainty is equal to the smallest digital increment, and no estimating may be performed. For example a stopwatch reading may be represented as t = 12.63 s ± .01 s.
Vernier Caliper
Another example of an instrument that requires no estimating is the Vernier caliper, which is used when greater precision is needed that what the meter stick can provide. The Vernier caliper is a device that employs an auxiliary (Vernier) scale which permits more accurate measurements of small fractions of a main scale division. The Vernier scale consists of equally spaced divisions which have the same total length as (n -1) main scale divisions. As an example, consider the caliper shown in Figure 2:
Figure 2
The main (stationary) scale is divided into millimeters. The Vernier (moveable) scale is 9 mm long and consists of 10 divisions. Each division on the Vernier scale, then, is 0.9 mm long or 0.1 mm shorter than a main scale division. Note that the first division on the Vernier scale is 0.1 mm from the first division on the main scale; the second Vernier division is 0.2 mm from the second main division scale division, etc. If the Vernier is adjusted so that the first divisions of the two scales are aligned, the distance between the jaws of the caliper is 0.1 mm. If the jaws are opened further, until the second divisions on the scales are aligned, the distance between the jaws will be 0.2 mm.
In general, the number of the Vernier line which is aligned with a line on the main scale is equal to the number of tenths of a millimeter between the caliper jaws. If the separation between the jaws is more than one millimeter, the reading of the Vernier is added to the reading of the main scale up to the zero point of the Vernier scale. The uncertainty of the Vernier is the same as the least count, which is 0.1mm. Figure 2, therefore, shows a reading of 7.5 mm ± .1 mm, or 0.0075 m ± .0001 m.
Micrometer Caliper
If one requires greater precision than can be obtained with a Vernier caliper, a micrometer caliper is often used. This instrument consists of an integral number of equally spaced divisions. Consider as an example the micrometer caliper shown in Figure 3.
Figure 3
The screw of this instrument has 2 threads per millimeter so that it advances .5 mm for each revolution. Since there are 50 divisions around the rim, the least count and the uncertainty of this instrument is 1/50 x 0.5 mm = 0.01 mm. In reading the rim of the instrument, one must determine whether the barrel is on its first or second revolution after a main scale division (one can determine this by simple inspection).
Metric Dial
Another useful measuring device is the metric dial (see Figure 3). The least count of the big needle is 0.01 mm. There are 100 increments in a full circle of the needle. The small needle counts the number of times the large needle has made a complete circle, so its least count is 1 mm. There is a set screw on the side of the dial that will allow one to zero the markers to the needle, then retighten.
Figure 4
You May Also Find These Documents Helpful
-
Data and Observations:Be sure that all measurements have the correct number of sig-figs and the correct units…
- 443 Words
- 2 Pages
Satisfactory Essays -
3 Measurements are to the correct number of significant figures Units included Answers are within the range of acceptable error Measurements finished within the prescribed time limit All measurements…
- 2395 Words
- 10 Pages
Good Essays -
When adding or subtracting quantities, the number of significant figures is rounded off to the same decimal place as the least precise piece of data. For example, when subtracting 23.385 cm from 18472.3952 cm, the answer will be rounded to the third decimal place, leaving our answer to be 18449.010 cm. It is important to note that defined quantities, such as the number of people at a table, or minutes in an hour,are not subjected to the rules of significant figures because they would not fall victim to erroneous measurements.…
- 624 Words
- 3 Pages
Good Essays -
________ misstatements are those where the auditor can determine the amount of the misstatement in the account.…
- 594 Words
- 3 Pages
Satisfactory Essays -
Significant figures: The number of digits you quote when you write down the value of a quantity that has been measured with a degree of uncertainty. The smaller the number of significant figures quoted for a value, the smaller the uncertainty in that value.…
- 429 Words
- 2 Pages
Satisfactory Essays -
3) Given the value calculated by the senior technician calculate your error and comment on this error in relation to the apparatus error of the method.…
- 1424 Words
- 5 Pages
Powerful Essays -
Task 5. Compute the percent error of the experimental value of g determined from the graph in part d. (Accepted value of g = 9.8 m/s2)…
- 260 Words
- 3 Pages
Satisfactory Essays -
Look at the calibration marks on your ruler to determine the degree of uncertainty and number of significant figures that can be made when measuring with a ruler.…
- 2629 Words
- 11 Pages
Good Essays -
If Patty wants to show the bank the maximum profit over the previous 2-year period, which costing method should she present?…
- 344 Words
- 2 Pages
Good Essays -
2). Make a table to record all of your data. Start measuring the diameter and circumferences of all your circular objects.…
- 391 Words
- 2 Pages
Good Essays -
Confidence Interval: A confidence interval is an indicator of a measurement's precision. It is also an indicator of how stable an estimate is, which is the measure of how close a measurement will be to the original estimate if an experiment is repeated.…
- 1058 Words
- 5 Pages
Better Essays -
Procedure: I used a measuring device such as a ruler, balance or volumetric flask to determine the values for each measurement.…
- 1411 Words
- 11 Pages
Satisfactory Essays -
The Uncertainty Reduction Theory deals with behavioral and cognitive aspects. Behavioral uncertainty is usually concerned with how to behave and following accepted conventions or protocols. On the other hand, Cognitive Uncertainty is concerned with who is this person and acquiring more information about the other. The general concept of ethics is related in a way that it also deals with morality and the principles relating to right and wrong conduct.…
- 224 Words
- 1 Page
Satisfactory Essays -
| x = 1n i=1naiWhere, 1. n refers to the total number of values. 2. ∑ refers to the addition of all values starting with the first value, denoted by i = 1, and ending off with the last value, denoted by n. 3. ai refers to the values in sequence from i = 1 to the nth term or last term.…
- 1699 Words
- 7 Pages
Better Essays -
When an alkali neutralizes an acid, a salt and water are formed. Aqueous hydrogen ions (H+(aq)) from the acid react with the hydroxide ions (OH- (aq)) from alkali, forming water. The identity of salt will depend on the nature of the acid and alkali used.…
- 1572 Words
- 7 Pages
Powerful Essays