Preview

Riboflavin Synthesis Lab Report

Good Essays
Open Document
Open Document
422 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Riboflavin Synthesis Lab Report
Based on the data from the riboflavin spectrum scan, the maximum absorbance wavelength for this compound is 446 nm. This was the point between 390 nm and 500 nm at which the absorbance value (0.72) was the highest. A blank tube that has the components of the solution being examined except for the compound of interest is then used in combination to provide an even more accurate reading. Then, by using Beer’s Law, the molar extinction coefficient for riboflavin was able to be calculated: 14,400 L/(moles*cm). This is also supported by the experimentally derived extinction coefficient from the standard curve equation (specifically the slope): 13,236.36. The slope of the standard curve is directly correlated with the extinction coefficient and in this experiment, is the same number, for the diameter of the cuvette being used was 1 cm. If this diameter was increased, then the slope of the standard curve would react by decreasing. These values are close, however, not the same. This could be due to a slightly less than perfect R2 value or pipette delivery variation when creating the dilution solutions. …show more content…
This is the point at which the spectrophotometer is the most sensitive to Riboflavin and can optimally detect its presence in the solution. This, in turn, provides the most accurate absorbance reading of the specific compound of interest. By using this value and the standard curve generated, the concentration of an unknown solution was able to be derived. Based on the results from this experiment it can be deduced that solution #36’s riboflavin concentration is approximately 0.000056 M. This value was quantitatively obtained by diluting the original solution by a factor of 5 and 10. Both dilution factors were tested in duplicate, producing results that are more likely to be

You May Also Find These Documents Helpful

  • Good Essays

    CHM130 Lab 6

    • 1556 Words
    • 6 Pages

    - The control here was that we used the same amount of each liquid, 10mL, to ensure that the experiment was giving us a reasonable comparison.…

    • 1556 Words
    • 6 Pages
    Good Essays
  • Good Essays

    CHEM 1252 Lab Report 5

    • 1825 Words
    • 21 Pages

    The absorbance of a small test tube filled with 4.0 mL of the 0.2 M NaOH solution and 1 drop of phenolphthalein was recorded every 5 seconds for 360 seconds. Then, ln(absorbance) and 1/(absorbance) were calculated for these values…

    • 1825 Words
    • 21 Pages
    Good Essays
  • Good Essays

    Experiment 3 prelab

    • 345 Words
    • 2 Pages

    The objectives of this lab include- illustrating the use of the spectrophotometer in chemical analysis, and generating a standard, or calibration curve, then using that curve to determine the value of an unknown substance. The spectrophotometer is one of the most powerful tools used in chemistry to find the concentration of substances in solution. It compares the colors of a known and an unknown solution, that comparison then leads to a quantitative estimate of the concentration of trace amounts of colored materials in that solution.…

    • 345 Words
    • 2 Pages
    Good Essays
  • Good Essays

    This could be due to the many assumptions being made in this experiment. We are assuming that there was no cross contamination between Solution A, B and water. Though it is highly likely that someone used the same pipette or measuring cylinder to measure chemicals as all three solutions were clear and colourless as notes in the overall observations. This could affect the content of each of the beakers as it would change the concentrations and/or volume of the solutions which was to be kept constant (refer to Table 1.1) greatly affecting the data obtained.…

    • 1146 Words
    • 5 Pages
    Good Essays
  • Good Essays

    Enzyme Lab Report

    • 492 Words
    • 2 Pages

    The first solution was tube 1 which was made with 1 mL of EDTA, 1 mL of CO, and 5 drops of CAT. Tube 2 was made with 1 mL of PTU, 1 mL of CO, and 5 drops of CAT. Tube 3 was made with 1mL of distilled water, 1 mL of CO, and 5 drops of CAT. After mixing each solution and putting a piece of Parafilm on Tube 1, we set tube 3 as our blank and then measured the change in absorption at 400nm of both tube 1 and tube 2 every two minutes for a total of twenty minutes.…

    • 492 Words
    • 2 Pages
    Good Essays
  • Better Essays

    Reactions were then incubated at 30°C for 10 minutes, and then placed on ice for 5 minutes. After incubation and cooling, the absorbances of the various samples were analyzed via spectrophotometry, and the final…

    • 2363 Words
    • 10 Pages
    Better Essays
  • Better Essays

    Pa Synthesis Lab Report

    • 1377 Words
    • 6 Pages

    5 mg of MT standard substance was weighed and dissolved in ultra-pure water to prepare the standard curve. The resulting solution was transferred to a 5 mL volumetric flask to prepare a 1 mg/mL DEX solution, which was then further diluted to prepare standard solutions of 5, 6.25, 10, 12.5, and 20 g/mL. The absorbance at 242 nm was measured using a UV-2600i spectrophotometer, and a standard curve was plotted with the concentration as the abscissa and absorbance as the ordinate. Next, the solution below and above the ultrafiltration membrane was collected, and the volume was…

    • 1377 Words
    • 6 Pages
    Better Essays
  • Better Essays

    Bio Spectroscopy Lab

    • 1273 Words
    • 6 Pages

    This investigation used spectroscopy to evaluate light absorption in different solutions. A spectrophotometer was used in the lab to determine these values. A spectrophotometer is an apparatus used to “measure the absorption of radiation in the visible and UV regions of the spectrum and allows precise at a particular wave length” (Jones et al., 2007). The amount of light absorbed by a substance is directly in relation to the concentration of the solute and also the wavelength moving through the solute (Jones et al., 2007). This is commonly referred to as Beer’s Law and can be expressed as A= εl [C]. Beer’s Law equation measures the absorbency of light, making it an effective measure as spectrophotometers give exact values for absorbency (Jones et al., 2007).…

    • 1273 Words
    • 6 Pages
    Better Essays
  • Good Essays

    The experiment of the controls 0.2, 0.3, 0.6 and 0.9 mg/ml protein yielded optical densities of .231, .329, .645 and .970, respectively. Figure 1 shows a positive correlation between absorbance and concentration; as the protein concentration increased, so did the absorbance.…

    • 354 Words
    • 2 Pages
    Good Essays
  • Powerful Essays

    MICRO BIOLOGY

    • 1277 Words
    • 6 Pages

    We used the dilution of 10-6 in the calculation to achieve the mean value against concentration as it was the most appropriate dilution…

    • 1277 Words
    • 6 Pages
    Powerful Essays
  • Powerful Essays

    Finding Density Lab

    • 1151 Words
    • 5 Pages

    was done according to the first but the volume was different, instead of 10 ml, 20 ml was measured.…

    • 1151 Words
    • 5 Pages
    Powerful Essays
  • Powerful Essays

    Although the hypothesis was accepted, there are still possibilities of errors while conducting this experiment. Rounding off the digits of the calculations and not using the appropriate amount of sig figs could have thrown the data off. The other limitations to look at is the possible limitation of the actual equipment. While conducting the experiment, the measurements could have been off or the containers the solutions were put in could have not been dried completely. The excess of water might have caused an error in the data…

    • 2144 Words
    • 9 Pages
    Powerful Essays
  • Good Essays

    If I want to target the peptide to different compartments, I would use specific signal sequences or uptake-targeting sequences encoded into my peptide. Signal sequences contain information that will target the peptide to a particular organelle. The information is usually encoded within amino acid sequences of the protein itself, usually within the 20-50 amino acids. Also, each organelle has specific set of receptors that bind only to the specific signal sequences, thus making sure of target specificity. Once the peptide containing the signal sequence has bound to the respective receptor, the peptide is transferred through the lipid bilayer through the translocation channel.…

    • 541 Words
    • 3 Pages
    Good Essays
  • Powerful Essays

    Gen 480

    • 1307 Words
    • 6 Pages

    The purpose of this paper is to reflect on my personal and professional growth during my Criminal Justice program of study at the University of Phoenix. The paper will reflect upon where I was before I began college at the University of Phoenix. Then I will evaluate the growth I experienced during my University of Phoenix program of study. To conclude, the paper will analyze the impact of completing the University of Phoenix bachelor's program on my current and future professional goals.…

    • 1307 Words
    • 6 Pages
    Powerful Essays
  • Good Essays

    Investor Ratio Analysis

    • 896 Words
    • 4 Pages

    There are various ratios that are designed to help investors who hold shares in a company to assess the returns on their investment. These are:…

    • 896 Words
    • 4 Pages
    Good Essays