Preview

2 Photon vs Confocal Microscopy

Good Essays
Open Document
Open Document
272 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
2 Photon vs Confocal Microscopy
Compare and contrast laser scanning confocal and multiphoton microscopy in 250-500 words.

Both confocal and two-photon (multi-photon) laser imaging can generate 3D images by capturing a stack of optical sections at different focal planes. In confocal microscopy, a detector pinhole which rejects fluorescence from off-focus locations is used; while, two- photon microscopy is capable of giving 3D contrast and resolution (comparable to confocal microscopy) without the necessity for the detector pinhole used in the confocal microscope. Two-photon microscopy doesn’t need the pinhole because the fluorophores are exclusively excited at the focus, where two photons meet each other and provide enough energy together for fluorophore excitation. This is important when we consider that the pinhole rejects signal photons emitting from the focus that are scattered on their way out of the tissue; thus, deep in tissue, confocal microscopy becomes unacceptably wasteful in terms of signal photons. Compensating for signal-loss with increased fluorescence excitation, confocal imaging leads to more phototoxicity and photobleaching. Therefore, experiments deeper in tissue benefit from two-photon excitation. The excitation wavelengths used in Two-photon microscopy (Infra-red) penetrate tissue better than the visible wavelengths used in one-photon microscopy. This improved penetration is due to both reduced scattering and reduced absorption by endogenous chromophores. Two-photon also causes less photodamage (photobleaching and phototoxicity), which used to be a limitation for the application of fluorescence microscopy to living systems. That’s because of low energy laser lights (Infra-red) application, and maximizing the probability of detecting a signal photon per excitation event (considering the fact that each excitation event has the risk of

You May Also Find These Documents Helpful

  • Good Essays

    Nt1310 Unit 3 Lab

    • 2577 Words
    • 11 Pages

    The bright field microscope is best known to students and is most likely to be found in a classroom. Visible light is focused through a specimen by a condenser lens, then is passed through two more lenses placed at both ends of a light-tight tube. The latter two lenses each magnify the image. Limitations to what can be seen in bright field microscopy are not so much related to magnification as they are to resolution, illumination, and contrast. Resolution can be improved using oil immersion lenses.…

    • 2577 Words
    • 11 Pages
    Good Essays
  • Good Essays

    Dye Lab for Chem 205

    • 2366 Words
    • 10 Pages

    One of such means termed fluorescence occurs when a substance can be induced into giving off light is through absorption of a photon through light or other means of radiations. During this process, electrons are excited from their standard state. The electrons jump from their highest unoccupied orbital (HUMO) into a certain level of the unoccupied orbital (LUMO). As the electrons loose energy, they fall back to a lower orbital, thus emitting light. 1, 2…

    • 2366 Words
    • 10 Pages
    Good Essays
  • Better Essays

    1. Length: A metric ruler is useful for measuring items of length. The ruler below measures in mm, indicated by the small mm near 0.…

    • 2331 Words
    • 10 Pages
    Better Essays
  • Good Essays

    Lab 1 Microscope

    • 516 Words
    • 3 Pages

    A. Ocular Lens- the lens on the top of the microscope that look into with 10x or 15x power. B. Body Tube- Connects eye piece to the objective lens C. Revolving Nosepiece- holds two or more objectives lenses and can be rotated easily to change power D. Objective lenses- 4x, 10x, 40x, and 100x powers E. Stage- The flat plate where the slides are placed for observation F. Diaphragm- Generally a five hold disc placed under the stage. Used to vary the intensity and size of the cone of light to see the slide. G. Illuminator- A light source, used to reflect light from an external light source up through the bottom of the stage. H. Coarse Focus Knob- Rough focus knob on the microscope used to move the objective lenses towards or away from the specimen. I. Fine Focus Knob- Knob used to fine tune the focus on the specimen, used after the coarse focus knob. J. Arm- Part of microscope that connects the tube to the base, used when carrying. K. Stage Clip- Clips on the stage used to hold the slide into place L. Base- bottom support of the microscope B.…

    • 516 Words
    • 3 Pages
    Good Essays
  • Satisfactory Essays

    Lab 3 Part 1

    • 830 Words
    • 9 Pages

    Two parts of the microscope that regulate the amount of light visible through the ocular are the:…

    • 830 Words
    • 9 Pages
    Satisfactory Essays
  • Better Essays

    BSP, S. (2010). How is EM different from light microscopy? Retrieved April 25, 2015, from http://bsp.med.harvard.edu/node/222…

    • 1286 Words
    • 4 Pages
    Better Essays
  • Good Essays

    Unit 45

    • 625 Words
    • 3 Pages

    Acoustic and ultrasonic microscopes, compound microscopes, and fluorescent and UV microscopes are available. Acoustic and ultrasonic microscopes use sound waves to create images of the sample. Compound microscopes use a single light path. These types of microscopes can have a single eyepiece (monocular) or a dual eyepiece (binocular). Compound microscopes have low depth perception but high resolution and magnification. Fluorescent microscopes and UV microscopes use high-energy and short-wavelength light to excite electrons, causing them to shift to higher orbits. When the electrons fall back to their original energy levels, they emit lower-energy and longer-wavelength light.…

    • 625 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Compound Light Lab

    • 942 Words
    • 4 Pages

    1. What is the difference between a simple light microscope and a compound light microscope?…

    • 942 Words
    • 4 Pages
    Good Essays
  • Good Essays

    2. I would also like to know how to properly and fully operate my vehicle so I can safely get…

    • 452 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Compound Light Microscope

    • 901 Words
    • 4 Pages

    Compound light microscope: is an optical instrument comprised of more than one lens and uses light to illuminate the object under study…

    • 901 Words
    • 4 Pages
    Good Essays
  • Good Essays

    The disadvantage of using a light microscope is its inability to magnify a specimen as great as an electron microscope. An electron microscope can magnify up to 500,000 times, whereas, a light microscope can only magnify an object up to 2000 times.…

    • 644 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Half Critical Lense

    • 324 Words
    • 2 Pages

    Pythagoras once said, “Choices are the hinges of destiny.” The decisions that people make will ultimately lead them to their future. Ray Bradbury and Ayn Rand illustrate this idea in Fahrenheit 451 and Anthem, respectively. Both authors portray this idea with the characters decisions in each novel.…

    • 324 Words
    • 2 Pages
    Good Essays
  • Powerful Essays

    Guided Photons

    • 8326 Words
    • 34 Pages

    A waveguide is a structure that guides waves, such as electromagnetic waves or sound waves. There are different types of waveguides for each type of wave. The original and most common meaning is a hollow conductive metal pipe used to carry high frequency radio waves, particularly microwaves. The geometry of a waveguide reflects its function. Slab waveguides confine energy to travel only in one dimension, fiber or channel waveguides for two dimensions. The frequency of the transmitted wave…

    • 8326 Words
    • 34 Pages
    Powerful Essays
  • Good Essays

    Hirase and colleagues from Rafael Yuste’s lab at Columbia University applied laser illumination to depolarize and excite single neurons in vitro. This work follows on the classic work of Fork who, in the early 1970s, demonstrated that laser illumination could produce excitation of molluscan neurons through a reversible, but unknown mechanism. Hirase et al., reporting in the Journal of Neurobiology, used modern two-photon techniques that enabled the laser light to be focused much more precisely than the technology used by Fork. They demonstrated that excitation of pyramidal neurons in brain slices from mouse visual cortex required the illumination to be applied tangential to the membrane of the cell, and that excitation was ineffective if the laser was focused below or within the cell.…

    • 699 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Nishiyama and his lab team experimented on the brains of adult mice. Surgery was performed as intraparenchymal injections, which contained a virus, were put in between the cerebellar vermal lobule VII and the paramedian lobule in the right cerebellar hemisphere. Three weeks after the injection was made, a cranial opening was made near the paramedian lobule and a coverslip was created right on top of the dura and the surgical cyanoacrylate. The researchers waited at least two weeks for the mice to recover, and an in vivo two-photon microscopy was performed by using a laser-scanning confocal microscope and external, non-descanned photomultiplier tubes. The parallel fibers were identified in raw z-stacks, and their changes in time were recorded. The size of appearance and disappearance of varicosities were noted at their appropriate times, and the mean was taken since there were many experimental factors that could change during each imaging session.…

    • 729 Words
    • 3 Pages
    Good Essays