Preview

Estimation Of LOD And LOQ Lab Analysis

Good Essays
Open Document
Open Document
953 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Estimation Of LOD And LOQ Lab Analysis
3.2. Estimation of LOD and LOQ using LMS and IRLS Limit of detection (LOD) according to Miller [51] is equal to YB+3 SB where YB is the value of the calculated intercept and SB is the Sy/x while limit of quantitation LOQ will be equal to YB+10 SB LOD and LOQ for each compound at each case were calculated. The LOD and LOQ after the robust regression line fitting of data were lower than those obtained before the treatment of data in both cases of linearity. As seen in Table 23, the LOD and LOQ were decreased from microgram range; 11.45 and 34.36 µg/mL to become in the nanogram range; 0.70 and 2.10 ng/mL, respectively, upon applying LMS for ∆F method in the ideal case of linearity. The effect is much more pronounced in non-ideal case of linearity. This indicated that the linearity ranges for …show more content…
The data of fluorimetric method, ∆F and F-ratio, treatment was done using OLS, LMS and IRLS in both cases of linearity. The values of recovery % were in the accepted ranges (98-102%) indicating the good accuracy of the method obtained after treatment of data, such effect is highly significant in the non-ideal case as they came closer to 100%. For example in Table 24, in non-ideal case, for ∆F-method, upon using LMS regression, the values of mean recovery% of concentration levels 20, 40 and 60 µg/mL increased from 91.20 % to 100.23%. In the same time, the RSD% for the different concentration levels used in the study showed a great decrease in their values upon applying both LMS and IRLS robust regression methods. This indicates that the proposed robust, non-parametric methods are superior and totally valid for the estimation of the unknown drug concentrations in the whole linearity range with acceptable accuracy and

You May Also Find These Documents Helpful

  • Satisfactory Essays

    Immediately, 100 µL of BEC-HRP conjugate diluted using PBS (1:15000) was added to the standard solution. The microtiter plate was incubated for another hour while shaking at 750 rpm. Consequently, another three cycled washing step was performed, 200 µL of the freshly prepared HRP substrate solution was added to each well. The HRP substrate solution25 was 21 ml citrate buffer (220 mM sodium citrate monobasic, pH 4.0) + 8.1µL H2O2 (30%) + 525 µLTMB solution (40 mM TMB, 8 mMtetrabutylammonium borohydride, in N,N-Dimethylacetamide). The microtiter plate was incubated for 30 minutes. Eventually, the color development was stopped by adding 100µL of 1 M H2SO4. And therefore, the absorbance was measured using plate spectrophotometer SpectraMax Plus384 (Molecular Devices, Ismaning, Germany) at 450 nm with 620 nm as a reference. Four parameter fit (4PL) was used for measuring Cocaine ELISA calibration curve; A (upper asymptote), B (slope at the inflection point), C (the inflection point on the calibration curve), and D (lower…

    • 309 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    ka lab report

    • 692 Words
    • 3 Pages

    We will be using the LoggerPro and LabPro in order to help us determine our data. The purpose of this experiment is to follow the changes of pH during the titration of an acid and a base in order to determine the of the weak acid, . is a constant for a given acid at a given temperature. In this experiment we determined the Ka using two different methods: 1) the measurement of the pH of a solution containing a known concentration of a weak acid, and 2) measurement of the pH at the half-neutralization point in the titration of a weak acid and a strong base.…

    • 692 Words
    • 3 Pages
    Good Essays
  • Good Essays

    In order to be able to relate absorbance readings at 405 nm to nmol p-nitrophenol/ml produced in the reaction mixtures at 25 OC and 35 oC a calibration curve must be set up, to cover the range 0-150 nmol/ml (0-150µM)…

    • 613 Words
    • 3 Pages
    Good Essays
  • Good Essays

    Purpose: To learn about the international system of units (SI), to become familiar with common lab equipment and techniques, to gain proficiency in determining volume, mass, length, and temperature of a variety of items using common laboratory measurement devices, to learn to combine units to determine density and concentration, and to use laboratory equipment to create serial dilutions and determine the density and concentration of each dilution.…

    • 1509 Words
    • 7 Pages
    Good Essays
  • Powerful Essays

    min . The calibration curve was linear over the range 0.1-5.0μg/ml. The method was also suitable for the…

    • 3534 Words
    • 26 Pages
    Powerful Essays
  • Good Essays

    Labpaq Scientific Method

    • 1167 Words
    • 5 Pages

    | Vertical tear: relatively even tear, frayed edges, very thin strands sticking outHorizontal Tear: rigid edges->less even, frayed edges, thin strands…

    • 1167 Words
    • 5 Pages
    Good Essays
  • Good Essays

    Labpaq Experiment 1

    • 979 Words
    • 4 Pages

    |The solution on the paper towel did not change in color even after exposure to bright light for 3 minutes.|…

    • 979 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    Lab 1 again

    • 768 Words
    • 3 Pages

    Figure 2: Absorbance readings of different concentrations of BSA. Individual data is presented of one group is presented (n=1). Line of best fit (y = 0.2091x - 0.215 R² = 0.9527) is presented using the trend line function in Excel.…

    • 768 Words
    • 3 Pages
    Satisfactory Essays
  • Powerful Essays

    As indicated by the diagrams above, both TLC plate tests went as expected and showed that the separation went very well. For the TLC plate in toluene, both the fluorene and fluorenone shifted, but the fluorenone shifted farther. This is expected since the polarity of toluene and fluorenone are more similar. However, since fluorene shifted to some degree as well, this indicates that fluorene is slightly soluble in toluene. Since the Rf values are 0.53 for fluorenone and 0.29, the test went as expected since fluorenone is more soluble in toluene. For the TLC plate in hexane, the fluorene moved up while the fluorenone stayed on the line, which was expected. The Rf values were 0.82 for fluorene and 0 for fluorenone. Overall the purpose was fulfilled as the TLC plates indicate that the two substances were separated well and also the IR spectra indicate that the experimental IR is very similar to the literature value.…

    • 1263 Words
    • 5 Pages
    Powerful Essays
  • Good Essays

    Gas Chromatography Essay

    • 921 Words
    • 4 Pages

    A reference sample is analysed and Rt values are compared to confirm the identity of a compound and to prepare a calibration curve for calculations of amounts, making GC a qualitative method.…

    • 921 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    Experiment 2

    • 532 Words
    • 3 Pages

    Figure 1. Sample calibration curve of standard Fluorescence solutions, average absorbance against concentration of standards. The concentration range of standard used was from 4.984x10-9 to 49.84ppm. The best fit line was determined by Regression analysis in Microsoft® Excel. The equation of best fit was y=0.1255 x +0.3353, the slope was 0.1255 V/ppm and the y intercept was 0.3353 V. The R2 value was 0.965.…

    • 532 Words
    • 3 Pages
    Satisfactory Essays
  • Good Essays

    HPLC

    • 981 Words
    • 4 Pages

    Use these data to plot calibration curves of peak area ratio vs. concentration for each analyte. Determine the regression equations. Attach the graphs to your report.…

    • 981 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    Gc-Mlc Lab

    • 70 Words
    • 1 Page

    Testing of GC-MS to know the chemical compound of remazol red both before and after the experiment, where the results of the GC-MS mass spectra and kromatogram form. Experimental results obtained by two types of kromatogram from GC that is before and after the experiment, kromatogram before the experiment shows more than one peak and shows the result of less smooth due to the substance used is remazol red…

    • 70 Words
    • 1 Page
    Satisfactory Essays
  • Good Essays

    Chromatography

    • 884 Words
    • 4 Pages

    8. The sample fine thin-layer chromatography plate, shown below, was prepared by spotting methyl red at R, sudan III at S, and bromocresol green at G. A single drop of each was placed on M. The plate was put in the developing solution until the solvent front reached 10 cm. Estimate the retention factor of R,S, and G, by measuring to the middle of the spot.…

    • 884 Words
    • 4 Pages
    Good Essays
  • Good Essays

    lab values

    • 831 Words
    • 4 Pages

    A magnesium test is used to measure the level of magnesium in the blood. Abnormal levels of magnesium are most frequently seen in conditions or diseases that cause impaired or excessive excretion of magnesium by the kidneys or that cause impaired absorption in the intestines. Magnesium levels may be checked as part of an evaluation of the severity of kidney problems and/or of uncontrolled diabetes and may help in the diagnosis of gastrointestinal disorders. Since a low magnesium level can, over time, cause persistently low calcium and potassium levels, it may be checked to help diagnose problems with calcium, potassium, phosphorus, and/or parathyroid hormone – another component of calcium regulation. Magnesium levels may be measured frequently to monitor the response to oral or IV magnesium supplements and may be ordered, along with calcium and phosphorus testing, to monitor calcium supplementation. Magnesium testing may be ordered as a follow up to chronically low levels of calcium and potassium. It also may be ordered when a person has symptoms that may be due to a magnesium deficiency, such as muscle weakness, twitching, cramping, confusion, cardiac arrhythmias, and seizures.…

    • 831 Words
    • 4 Pages
    Good Essays