Preview

Photosynthesis Lab Report

Good Essays
Open Document
Open Document
903 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Photosynthesis Lab Report
The majority of the experimentation in this project relied on the idea that there is a relationship between concentration and absorbance of a solution. This is easy to understand if you can visualize the particles in the solution; and this is especially true if the particles in solution portray a distinct color. The more particles there are in the solution, the more they will obstruct the path of the light. This occurs the most at the analytical wavelength, where the most light is absorbed by the particles. When light of this wavelength hits the particles, it is absorbed and there is no chance that it will reach the sensor in the spectrophotometer. However, if a solution is analyzed with a wavelength corresponding to the color of the particles, …show more content…

This solution had an analytical wavelength of 380 nm, because that is the wavelength at which the complex absorbs the most light; allowing the lowest amount to reach the sensor. The source of phosphate for making our standard curve was potassium hydrogen phosphate, and this wavelength allowed us to target and analyze only the phosphate-AVM complex while yielding the least impact from the potassium ions. As for the colas, they are made with so many ingredients that it was very important to have a good analytical wavelength. In addition to phosphate, there is caffeine, sugar, corn syrup, vanilla, and many other ingredients in colas (Field); all of which could have an effect on a percent transmittance reading from the Spec 20 at a poorly selected wavelength. Using the analytical wavelength minimizes the effect that other factors in the solution have in the reading, which in turn minimizes error in the …show more content…

The results of the food coloring exercise are seen in Figure 1 of the group report. The red food coloring displayed the highest percent transmittance in the long wavelengths closest to red light, which was expected. With the same logic, it was expected that the blue food coloring would have the highest percent transmittance in the higher energy wavelengths associated with blue light. In fact, the trend lines for the red and blue food coloring look almost like opposites. As for the potassium permanganate, we were interested in the wavelength that resulted in the lowest percent transmittance. This wavelength, which was 550 nm, was our analytical wavelength. This wavelength yields the most accurate results when analyzing potassium permanganate and also provides for the least error from other substances in the solution. Choosing the proper analytical wavelength for phosphate was very important for testing the colas since there are so many different ingredients that go into

You May Also Find These Documents Helpful

  • Satisfactory Essays

    Immediately, 100 µL of BEC-HRP conjugate diluted using PBS (1:15000) was added to the standard solution. The microtiter plate was incubated for another hour while shaking at 750 rpm. Consequently, another three cycled washing step was performed, 200 µL of the freshly prepared HRP substrate solution was added to each well. The HRP substrate solution25 was 21 ml citrate buffer (220 mM sodium citrate monobasic, pH 4.0) + 8.1µL H2O2 (30%) + 525 µLTMB solution (40 mM TMB, 8 mMtetrabutylammonium borohydride, in N,N-Dimethylacetamide). The microtiter plate was incubated for 30 minutes. Eventually, the color development was stopped by adding 100µL of 1 M H2SO4. And therefore, the absorbance was measured using plate spectrophotometer SpectraMax Plus384 (Molecular Devices, Ismaning, Germany) at 450 nm with 620 nm as a reference. Four parameter fit (4PL) was used for measuring Cocaine ELISA calibration curve; A (upper asymptote), B (slope at the inflection point), C (the inflection point on the calibration curve), and D (lower…

    • 309 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    CHEM 1252 Lab Report 5

    • 1825 Words
    • 21 Pages

    For the first part of the experiment, 4.0 mL of the 0.2 M NaOH solution were poured into a small test tube. The test tube was wiped and placed into the sample holder of the spectrophotometer and the transmittance was set to 100.0 for calibration. The test tube was removed from the sample holder and 1 drop of phenolphthalein was added and inverted several times to mix. With the Spec-20 set to absorbance mode, the test tube was quickly placed into the sample holder and the absorbance was recorded every 5 seconds for 360 seconds. For the second part of the experiment, the same calibration steps were repeated and then, 2.0 mL of the 0.2 M NaOH and 2.0 mL of the 0.2 M NaCl were added to a small test tube. After this step, 1 drop of phenolphthalein was added and inverted several times to mix. With the Spec-20 set to absorbance mode, the test tube was quickly placed into the sample holder and the absorbance was recorded every 5 seconds for 360 seconds. After each part of the experiment, the test tube was removed from the sample holder and its contents were placed in the waste.…

    • 1825 Words
    • 21 Pages
    Good Essays
  • Satisfactory Essays

    In our experiment, there was a source of error because in the Concentration vs. Absorbance graph’s line of best fit, there was a y-intercept, but there isn’t suppose to be one since the concentration is 0, the absorbance should be 0. There are no particles in the solution, distilled water, so nothing should be absorbed. This error could have been caused by the leaving the diluted solution out too long, so the water in the solution could have evaporated, leaving only the blue dye. It is not a diluted solution if there is just dye left in the beaker because it would go be a higher concentration than the actual solution’s concentration…

    • 133 Words
    • 1 Page
    Satisfactory Essays
  • Satisfactory Essays

    Chem Lab report

    • 452 Words
    • 2 Pages

    Materials: Beakers, 50-mL, 2-3 ; Test tubes (13 x 100)mm 3-8 ; FD&C Blue 1 stock solution, 50-mL ; Lens tissues ; Pipets, serological, 10-mL ; Sample of consumer sports drink, 10-mL ; Spectrophotometer ; Test tube rack…

    • 452 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Experiment 3 prelab

    • 345 Words
    • 2 Pages

    The objectives of this lab include- illustrating the use of the spectrophotometer in chemical analysis, and generating a standard, or calibration curve, then using that curve to determine the value of an unknown substance. The spectrophotometer is one of the most powerful tools used in chemistry to find the concentration of substances in solution. It compares the colors of a known and an unknown solution, that comparison then leads to a quantitative estimate of the concentration of trace amounts of colored materials in that solution.…

    • 345 Words
    • 2 Pages
    Good Essays
  • Satisfactory Essays

    We then measured the absorbance of samples of grape soda, diet cherry, strawberry crush, and big red soda using the same spectrophotometer set at the same wavelength. Our standard curve correlating concentrations to absorbance values only went to an absorbance of 1 nm. So, for three of the four drinks, it was necessary to dilute the substance to get an absorbance rate within our standard curve. After the absorbance rates of these drinks were found, we used the standard curve to determine the concentration of red dye #40 in each drink. To do this, we looked to see which concentration matched with each measured…

    • 342 Words
    • 2 Pages
    Satisfactory Essays
  • Satisfactory Essays

    By using a spectrophotometer with the unknown solution, a percent transmittance of 56 appeared. The concentration of the unknown solution was then easily plotted on a graph to show the corresponding data.…

    • 258 Words
    • 2 Pages
    Satisfactory Essays
  • Satisfactory Essays

    Photosynthesis Lab Report

    • 566 Words
    • 3 Pages

    Photosynthesis is a process to which some autotrophs such as plants produce their own food. It has two stages or reactions light-dependent and light-dependent reactions. The light dependent reactions are the first stage, where energy from sunlight is captures in Photosystem 2 and then 1, the electrons generated in Photosystem 1 then moves along the electron transport chain. The moving of electrons causes a hydrogen ion gradient that is used in the final step to produce ATP, by the ATP Synthase. The equation for this reaction is 6CO2 + 6H2O  C6H12O6+6O2 or, carbon dioxide +water sugars + oxygen . This reaction takes place on the thylakoid membrane of the chloroplast and uses energy from sunlight to produce compounds like ATP and NADPH. The second stage of photosynthesis is the Calvin Cycle, which takes place in the stoma of the chloroplast. This stage is light-independent or it does not need energy from sunlight. During this reaction ATP and NADPH from Stage 1are used to produce high energy sugars. The reaction is 3CO2 + 6NADPH + 5H2O + 9ATP  G3P + 2H+ + 6NADP+ + 9ADP + 8Pi . Additionally, the three limiting factors of photosynthesis are the light intensity, carbon dioxide concentration, and the temperature. Lastly, the items used in this particular experiment were baking soda to provide CO2, a light to excite the electrons in stage 1, distilled…

    • 566 Words
    • 3 Pages
    Satisfactory Essays
  • Powerful Essays

    urine when the wavelength of UV detection was 245 nm with acetanilide as the internal standard.…

    • 3534 Words
    • 26 Pages
    Powerful Essays
  • Satisfactory Essays

    The absorption of all wavelengths outside approximately the 510 nm to just over the 625 nm range and the reflection within that wavelength range produces the effect the eyes see the brain interprets that grass is green.…

    • 592 Words
    • 3 Pages
    Satisfactory Essays
  • Powerful Essays

    Penny Lab Report

    • 2257 Words
    • 10 Pages

    Introduction: In the spectrophotometric analysis of a penny lab, we will be conducting three experiments. The purpose of the first experiment is to find the maximum absorbance on the spectrometer to obtain the optimum wavelength. Once we have gathered that information, we will use the results in the other two experiments. The purpose of…

    • 2257 Words
    • 10 Pages
    Powerful Essays
  • Better Essays

    Coordination Complex Lab

    • 716 Words
    • 3 Pages

    A sample of copper sulfate pentahydrate (6.285g) was weighed out. The copper sulfate pentahydrate was dissolved in water (11.99mL) in a 250mL beaker. The solution was heated on a hot plate to 90 degrees Celsius. A sample of potassium oxalate monohydrate (10.006g) was dissolved in water (50.0mL).…

    • 716 Words
    • 3 Pages
    Better Essays
  • Good Essays

    In order to complete the objective, the group used a Carolina Digital Spectrophotometer, unit M 201-1, which measures the quantity of light that solutions absorb. The spectrophotometer consists of a light source that shines through a prism which then shines through a…

    • 1159 Words
    • 5 Pages
    Good Essays
  • Better Essays

    Bio Spectroscopy Lab

    • 1273 Words
    • 6 Pages

    This investigation used spectroscopy to evaluate light absorption in different solutions. A spectrophotometer was used in the lab to determine these values. A spectrophotometer is an apparatus used to “measure the absorption of radiation in the visible and UV regions of the spectrum and allows precise at a particular wave length” (Jones et al., 2007). The amount of light absorbed by a substance is directly in relation to the concentration of the solute and also the wavelength moving through the solute (Jones et al., 2007). This is commonly referred to as Beer’s Law and can be expressed as A= εl [C]. Beer’s Law equation measures the absorbency of light, making it an effective measure as spectrophotometers give exact values for absorbency (Jones et al., 2007).…

    • 1273 Words
    • 6 Pages
    Better Essays
  • Better Essays

    A spectrophotometer is commonly used for the measurement of transmittance or reflectance of solutions, transparent or opaque solids, such as polished glass, or gases. However they can also be designed to measure the diffusivity on any of the listed light ranges that usually cover around 200nm - 2500nm using different controls and calibrations. Within these ranges of light, calibrations are needed on the machine using standards that vary in type depending on the wavelength of the photometric determination.…

    • 1116 Words
    • 5 Pages
    Better Essays