Methods and materials:
In this lab we used a spectrophotometer, a test tube filled with water to be used as a blank, six test tubes with different concentrations of salicylic acid ranging from 0 mg/dL to 5 mg/dL and one test tube with an unknown concentration of salicylic acid.
We set the spectrophotometer to a wavelength of 540 nm. We adjusted the transmittance to 0%. Next we placed the test tube with the water into the spectrophotometer and adjusted the transmittance dial to 100%. We removed the test tube containing the water and replaced it with the first concentration of 0 mg/dL. We recorded the absorbance and repeated the test for a total of 5 readings. We did this for each concentration. We then calculated the average of each concentration’s absorbance readings and plotted the averages onto the graph. Once we had gathered the data for our known concentrations we then repeated the procedure for our unknown concentration. We again took the average and plotted that on the absorbance curve to determine the concentration of the unknown.
Observations and Data: Calibration Standard | Absorbance Reading 1 | Absorbance Reading 2 | Absorbance Reading 3 | Absorbance Reading 4 | Absorbance Reading 5 | AverageAbsorbance Reading | 0 mg/dL | 0.007 | 0.010 | 0.007 | 0.005 | 0.006 | 0.007 | 0.5 mg/dL | 0.032 | 0.036 | 0.037 | 0.041 | 0.038 | 0.037 | 1.5 mg/dL | 0.098 | 0.100 | 0.098 | 0.099 | 0.096 | 0.098 | 2.5 mg/dL | 0.150 | 0.149 | 0.153 | 0.154 | 0.150 | 0.151 | 3.5 mg/dL | 0.234 | 0.239 | 0.237 | 0.250 | 0.229 | 0.238 | 5.0 mg/dL | 0.286 | 0.287 | 0.288 | 0.292 | 0.291 | 0.289 | Unknown A | 0.241 | 0.238