Preview

Direct Determination Of Saliva Proteins Lab Report

Good Essays
Open Document
Open Document
510 Words
Grammar
Grammar
Plagiarism
Plagiarism
Writing
Writing
Score
Score
Direct Determination Of Saliva Proteins Lab Report
2.4.4. Direct determination of saliva proteins
Protein contaminated with nucleic acids absorbed the light at wavelength 280 nm and it absorbs much strongly at wavelength 205 nm when it is free from nucleic acids. The UV-visible spectrophotometer was used in determination of saliva proteins (Figure 2.2). Cold trichloroacetic acid (10 % w/v ) was added to the sample, centrifuged for 10 minutes to precipitate protein. The absorbance of a known volume ( 50 µl) of sample diluted with 4 milliliters distil water was measured at wavelengths 205 nm , 260 and 280 nm for each subject and patient . The concentrations of contaminated and uncontaminated proteins were determined by using the following equations (Layne, 1957; Scopes, 1974; Stoscheck, 1990):

You May Also Find These Documents Helpful

  • Good Essays

    CHEM 1252 Lab Report 5

    • 1825 Words
    • 21 Pages

    For the first part of the experiment, 4.0 mL of the 0.2 M NaOH solution were poured into a small test tube. The test tube was wiped and placed into the sample holder of the spectrophotometer and the transmittance was set to 100.0 for calibration. The test tube was removed from the sample holder and 1 drop of phenolphthalein was added and inverted several times to mix. With the Spec-20 set to absorbance mode, the test tube was quickly placed into the sample holder and the absorbance was recorded every 5 seconds for 360 seconds. For the second part of the experiment, the same calibration steps were repeated and then, 2.0 mL of the 0.2 M NaOH and 2.0 mL of the 0.2 M NaCl were added to a small test tube. After this step, 1 drop of phenolphthalein was added and inverted several times to mix. With the Spec-20 set to absorbance mode, the test tube was quickly placed into the sample holder and the absorbance was recorded every 5 seconds for 360 seconds. After each part of the experiment, the test tube was removed from the sample holder and its contents were placed in the waste.…

    • 1825 Words
    • 21 Pages
    Good Essays
  • Good Essays

    Lab Report

    • 1398 Words
    • 6 Pages

    The homogenates provided were made by homogenizing tissues in a sucrose phosphate buffer in a 1:20 ratio. The protein concentration in bovine cells was measured by diluting the homogenate with a 1:5 ratio; 50 microliters of homogenate and 200 microliters of water. Then 5 known protein concentration samples which were 0.4, 0.8, 1.2, 1.6, 2.0 mg/ml of bovine serum were used to determine absorbance with a spectrophotometer. Two additional samples were made; one was blank and the other was for the specific homogenate sample. Then 3 microliters of bradford assay reagent, which indicates the amount of protein present by color, was added to all samples. The spectrophotometer was zeroed at 595 nm. A standard curve was made with the different absorbencies and concentrations. After the linear equation was formed, the unknown sample concentration was determined using the standard curve equation. A Gel Electrophoresis was used to perform a qualitative analysis. The use of 5 microliters of the homogenate was heated to 80 degrees Celsius. Then the homogenate was transferred to a 2-microliter-protein gel sample buffer. Samples loaded on to the gel was run at 100 v and stained with comassie blue; observations were made next lab. (Clendening 2014)…

    • 1398 Words
    • 6 Pages
    Good Essays
  • Satisfactory Essays

    In order to create a standard curve we created 5 dilutions of solutions containing red dye #40. Concentrations of .666 ug/mL, 1.667ug/mL, 3.333ug/mL, 13.333ug/mL, and 20.000 were measured in a spectrophotometer set to a wavelength of 504 nm to find the amount of light they absorbed. Using Microsoft Excel, we entered the absorbance values as the Y coordinates and the concentrations as the X coordinates to make a line of best fit and that was our standard curve.…

    • 342 Words
    • 2 Pages
    Satisfactory Essays
  • Good Essays

    Tube 2 Lab Report

    • 603 Words
    • 3 Pages

    As indicated by the figure, the high concentration correlated to having the highest amount of absorbance (1.006 at 300 seconds). This was followed by the medium-high concentration (0.555) and medium concentration (0.540). It can be noted that the medium concentration started off with a higher absorbance than the medium-high concentration, but the medium-high concentration had a faster increase of absorbance over time. Thus, surpassing the absorbance of the medium concentration from 270 and 300 seconds. The low concentration had the lowest amount of absorbance, with a final absorbance rate of 0.204, and did not substantially increase over the period of…

    • 603 Words
    • 3 Pages
    Good Essays
  • Satisfactory Essays

    Bradford Protein Asssay

    • 393 Words
    • 2 Pages

    UV absorption can determine protein concentration, most Proteins, absorb ultraviolet light relatively strongly. It is the amino acids in the proteins that absorb the UV light. The strong absorbance of UV light by protein allows for rapid analysis of protein samples. The optical absorbance of protein is measured at 280 nm. At this wavelength, the absorbance of protein is largely due to the amino acids tryptophan, tyrosine and cysteine .The…

    • 393 Words
    • 2 Pages
    Satisfactory Essays
  • Powerful Essays

    urine when the wavelength of UV detection was 245 nm with acetanilide as the internal standard.…

    • 3534 Words
    • 26 Pages
    Powerful Essays
  • Good Essays

    In order to complete the objective, the group used a Carolina Digital Spectrophotometer, unit M 201-1, which measures the quantity of light that solutions absorb. The spectrophotometer consists of a light source that shines through a prism which then shines through a…

    • 1159 Words
    • 5 Pages
    Good Essays
  • Better Essays

    Bio Spectroscopy Lab

    • 1273 Words
    • 6 Pages

    This investigation used spectroscopy to evaluate light absorption in different solutions. A spectrophotometer was used in the lab to determine these values. A spectrophotometer is an apparatus used to “measure the absorption of radiation in the visible and UV regions of the spectrum and allows precise at a particular wave length” (Jones et al., 2007). The amount of light absorbed by a substance is directly in relation to the concentration of the solute and also the wavelength moving through the solute (Jones et al., 2007). This is commonly referred to as Beer’s Law and can be expressed as A= εl [C]. Beer’s Law equation measures the absorbency of light, making it an effective measure as spectrophotometers give exact values for absorbency (Jones et al., 2007).…

    • 1273 Words
    • 6 Pages
    Better Essays
  • Satisfactory Essays

    Lab 1 again

    • 768 Words
    • 3 Pages

    Figure 2: Absorbance readings of different concentrations of BSA. Individual data is presented of one group is presented (n=1). Line of best fit (y = 0.2091x - 0.215 R² = 0.9527) is presented using the trend line function in Excel.…

    • 768 Words
    • 3 Pages
    Satisfactory Essays
  • Good Essays

    Ninhydrin Lab

    • 910 Words
    • 4 Pages

    Yes, if the absorbance of the sample appeared to be outside the linear range, it is still possible to determine the protein concentration of the solution from the standard curve by diluting the solution. Dilution itself is a process of lessening a solute’s concentration in a solution. In this experiment, dilution can be done by reducing the amount of protein solution from 1.0 ml to 0.2 ml. Without the process of dilution, the absorbance will end up being an extrapolated value causing the result to be unreliable.…

    • 910 Words
    • 4 Pages
    Good Essays
  • Good Essays

    One gram of Tough Guy protein powder was added to 100 ml of water. 50 µl of the sample was added to respective test tubes: 0.2, 0.3, 0.6 and 0.9 mg/ml protein. 2.5 ml of diluted Bio-Rad assay solution was pipetted into each test tube. Each test tube was covered with Parafilm and placed over approximately 6-8 seconds. The solutions were placed at room temperature for one hour. Optical density was measured using a spectrophotometer at 595 nm. The relationship between the absorbance and concentration of the reaction was recorded on an Excel spreadsheet.…

    • 354 Words
    • 2 Pages
    Good Essays
  • Good Essays

    Proteins Biuret Test Results. Figure 1 is showing both graphs of results from the spectrophotometer. Figure 1A is the results without the unknown plotted. By using the best fit line and equation the unknown substance was found to have .75 absorbance and 11.5 mg/mL protein concentration. Figure 1B shows the newly found data of the unknown plotted with X.…

    • 883 Words
    • 4 Pages
    Good Essays
  • Good Essays

    Spinach Lab

    • 767 Words
    • 4 Pages

    First, the spectrophotometer was set to a wavelength of 605 nm. Then15 mL of pre-prepared chloroplast membrane suspension was obtained, and kept cold in a beaker containing ice to keep the chloroplast fragments from degrading. Then DCIP was retrieved in a foil wrapped test tube to keep any form of light out. Also, 15 mL of distilled water was obtained along with 7 test tubes; labeled blank and numbers 1-6. A blank was prepared by combining 5 mL of distilled water and 2.5 ml of chloroplast membrane suspension, mixing it well and producing a consistent green color. Tube 1 was set as a control for experiment three by combining .5 mL of DCIP and 2.5 mL of water. The absorbance of the first test tube was recorded initially at 0 minutes, and then placed under a light for one minute. Then the absorbance was tested again. and this process was repeated until there were five total readings. The second tube was used as a positive control against white light. The solution was produced by combining .5 mL of DCIP with 2.5 mL of CMS. Like the blank, this tube was screened for its original absorbance, and then exposed to white light for one minute until five readings were obtained. The third test tube was a negative control with no exposure to light. The tube was wrapped in a foiling casing and cap, and contained the same solution as test tube 2, .5 mL of DCIP and 2.5 mL of CMS. The…

    • 767 Words
    • 4 Pages
    Good Essays
  • Satisfactory Essays

    Hypothesis: The percentage of chewing gum that is sugar for bubble gum is 30%, 20% for juicy fruit and 25% for stride gum.…

    • 348 Words
    • 2 Pages
    Satisfactory Essays
  • Better Essays

    Protein Quantification

    • 1458 Words
    • 6 Pages

    Protein concentration analysis is primarily done through an accepted form commonly referred to as the Bradford Protein Assay. The main purpose of this experiment was to observe and record the various protein samples’ absorbency values through the calibrated readings of a spectrophotometer (595 nm calibration). Using the standard curve equation (y = 1.6147x + 0.0968) derived from determining the regression equation of different protein concentrations of BSA, an accurate form of protein concentrations were obtained for three different given samples, including an unknown. A notable empirical observation found in this experiment is the gradual increase in dark color change (i.e., blue) as the concentration of a protein increased with the inverse proportion of water per reagent added.…

    • 1458 Words
    • 6 Pages
    Better Essays